Eurosafe 170
Euronda
Automatisch erfasst von euronda.ro
Sie sind der Hersteller?Daten liefern
Herstellertext auf Englisch
Eurosafe 170 thermodisinfector
The Eurosafe 170 thermodisinfector meets the needs of clinics with a high number of patients. Efficiency and longevity are ensured by the injection washing and drying system and a water filtering system that preserves pump performance over time.
The Eurosafe 170 allows the user to choose 3 from 20 preset cycles by matching them to the buttons under the 3.5″ colour display, making them easy to select. It also allows the creation of customised cycles based on the specific needs of the material to be treated, adjusting duration and washing temperature.
The Eurosafe 170 is equipped with an integrated softener that allows the use of mains water and monitors the correct regeneration of the resins, giving a warning in the event of salt deficiency. The automatic dosing unit calibrates and injects the two detergent liquids, positioned outside the thermodisinfector, autonomously and without waste.
The thermodisinfector, at the end of each cycle, creates a file that can be transferred to a PC via USB, which reports all process data (time, temperature, cycle number), while also validating the load. In addition, thanks to optional accessories, data can be transferred via Ethernet connection or printed out thanks to the thermal printer.
Made in Italy with durable AISI 316L stainless steel.
Automated washing
Eurosafe 170 meets the requirements of EN ISO 15883 and allows the control of washing parameters (time, detergent concentration and temperature) during all phases. The effectiveness of thermal disinfection is measured by the A0 value, which can be achieved with different combinations of temperature and time: the higher the temperature, the less time is needed. The standard requires a minimum A0 of 600 for thermodisinfection of surgical instruments to ensure the elimination of bacteria, mycobacteria, fungi and viruses. Eurosafe 170 allows the A0 parameter to be varied to meet different needs and local regulations.
Pre-washing
This is carried out using cold water with the aim of removing organic substances and the most superficial and coarse dirt.
Washing
This is performed at temperatures between 50 and 65 °C, adding an enzymatic detergent to ensure perfect cleaning of the instruments.
Rinsing
This is done in order to remove any detergent residue from the load used in the previous phase.
Thermodisinfection
This takes place through the combined action of hot water at an adjustable temperature of up to 95°C and a variable time of 1 to 10 minutes. During the heating phase, a second chemical agent is injected into the chamber, which prevents corrosion of the load and the forming of deposits, and also facilitates the subsequent drying phase.
Drying
The final stage of the process is achieved through effective air distribution within the chamber to obtain, when the device is opened, perfectly dry material ready for packaging. In addition, it is possible to create customised cycles with different combinations of time and temperature. It is also possible to set a value higher and well above the standard, and monitor this parameter in real time via the display. In addition, a convenient button allows the forced-air drying system to be switched off, thus reducing energy consumption.
Load capacity
A comprehensive range of accessories including loading trolleys, inserts, baskets and adapters simplifies the operator's work and enables excellent cleaning, thermodisinfection and drying performance, maximising loading capacity and minimising waste. The available accessories meet the needs of all clinics working in the dental, maxillofacial surgery, oral surgery, orthodontics and general medicine fields.
Traceability
They can also be transferred to a PC via Ethernet connection or printed on thermal paper using the optional accessories. E-Memory is the software for the complete archiving, management and traceability of all sterilization cycles and data directly on a PC. This application can automatically, semi-automatically or manually capture data from the sterilization process and save it directly to the computer. This data can be easily organised, searched, printed and emailed.
Produktinformationen
| Hersteller | Euronda |
|---|---|
| Gewicht | 90 Kg |
| Anzeige | 3.5″ colour display |
| Material | AISI 316L stainless steel |
| Abmessungen | 60 x 84 x 63 cm (WxHxD) |
| Konnektivität | Included: USB and E-Memory software, Optional: Datalogger |
| Stromversorgung | Single-phase 220 V / 50Hz, Three-phase 380 V / 50Hz |
| Klassifizierung | Medizinprodukt Klasse IIb |
| Kammervolumen | 170 L |
| Power Consumption | 2700 W (220 V) – 5600 W (380 V) |