Hardware kit for validation and daily testing of VHP H₂O₂ plasma sterilizer penetration. Designed to simulate worst-case sterilization scenarios with lumens up to 4 meters in bidirectional configuration and 1.4 meters one-sided. Includes 1 mm diameter lumens and cavity for chemical indicator placement. Used to verify effective H₂O₂ vapor penetration into complex device geometries during daily confidence testing and formal validation (IQ/OQ/PQ). Compliant with ISO 11140, ISO 14937, and AAMI ST58 standards. Kit has no predefined shelf life when used according to operation manual and undamaged.

Produktinformationen

HerstellerTuttnauer Europe BV
HaltbarkeitNo predefined shelf life when undamaged
NormenkonformitätISO 11140, ISO 14937, AAMI ST58