Mirror Handle | ErgoSteel 11105

ErgoDenta ApS · #11105

Collecté automatiquement depuis ergodenta.dk

Vous êtes le fabricant ?Transmettez vos données

Texte du fabricant en anglais

The Mirror Handle is a lightweight, ergonomic cone-socket mirror handle. It threads onto any ErgoDenta screw-in mirror head, so you choose the head and the handle weight, colour and grip that suit you.

  • Cone-socket design — swap or replace the mirror head without changing the handle
  • Solid/hollow steel body for maximum tactile feedback and durability
  • Fully autoclavable and reusable

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.

  • Step 1 Pre-Cleaning: Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
  • Step 2 Automated Cleaning: Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
  • Step 3 Inspect & Maintain: Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
  • Step 4 Package: Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
  • Step 5 Sterilize: Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
  • Step 6 Store: Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.

Informations produit

FabricantErgoDenta ApS
TypeCone-socket mirror handle
StérilisationEntièrement autoclavable